Flow Cytometry Core @ CHOP

Tools | ImageStream guidelines

How to plan an imaging flow cytometry experiment, prepare your cells and place your fluorochromes.

The ImageStream takes an image of every cell it measures. This means sample preparation affects more than signal strength. It also affects where the signal sits in the cell, and that is usually what you want to measure. Use this guide when you plan your experiment, before you book time. Rev. 17NOV2026

Required: the Data Acquisition Form Fill it in before your reservation. At the end, click Copy link and paste the link into your iLab request. We use it to set up the instrument for your experiment. https://www.cytochop.org/guidelines/imagestream-request.html
Open the Data Acquisition Form →

The instrument at a glance

The ImageStreamX Mk II is an imaging flow cytometer. Cells flow in single file past the lasers, as in a flow cytometer, but instead of one number per channel it records an image of each cell in every channel (brightfield, side scatter and up to 10 fluorescence channels), thousands of cells per second. You can then measure not only how bright a cell is, but where the signal is: on the membrane or in the nucleus, in spots, co-localized with another marker, or how the cell is shaped.

  • Lasers: 405, 488, 561 and 642 nm for excitation. A 785 nm laser is used only for side scatter.
  • Objectives: 20×, 40×, 60×. The largest cell or particle each can image is about 120 µm at 20×, 60 µm at 40× and 40 µm at 60×.
  • Cameras and channels: 12 channels on two cameras. Camera 1 (Ch1–Ch6) collects dyes excited at 488 and 561 nm. Camera 2 (Ch7–Ch12) collects dyes excited at 405 and 642 nm.
  • Brightfield: uses one channel on each camera, chosen as a pair: Ch1 + Ch9 (the default), Ch2 + Ch8, Ch4 + Ch10, Ch5 + Ch11 or Ch6 + Ch12. It is used to focus and to align the two cameras’ images, so it is normally on. The remaining channels carry fluorescence.
  • Side scatter is optional. It uses one channel, Ch6 or Ch12.

Add your fluorochromes below to see which channel each one lands in. This is the ImageStream card of the Instrument Selector. It chooses the lasers, the brightfield pair and the side scatter channel that suit your panel, and you can change them.

To see where your fluorochromes land on the ImageStream, use the Instrument Selector at https://www.cytochop.org/instruments/panel.html.

Choosing fluorochromes

  • Use the brightest dye for the target with the fewest copies per cell. Aim for at least a full log shift over unstained cells, measured by flow.
  • Balance dyes that share a laser so they are within about a log of each other. A very bright dye saturates the camera at settings that a dim dye needs.
  • Similar emission, different cameras: two dyes with similar emission can be used together if their lasers are on different cameras. For example, with the default brightfield pair, PE-Dazzle 594 (561 nm) lands in Ch4 and BV605 (405 nm) lands in Ch10. Two dyes with similar emission excited by lasers on the same camera (405 + 642, or 488 + 561) land in the same channel and can’t be separated.
  • For nuclear counterstains, DAPI (on 405) and DRAQ5 (on 642) both work well. Choose one that doesn’t take a channel you need.

Preventing antibody capping

Antibodies that cross-link surface receptors can cause the receptors to cluster (“cap”) or be pulled into the cell within minutes at room temperature or 37 °C. In a flow cytometer this goes unnoticed. On the ImageStream it shows up as a localization pattern that has nothing to do with your biology. This is a common reason ImageStream data can’t be interpreted.
  • Stain on ice, using buffers pre-chilled to 4 °C.
  • Add 0.1% sodium azide to staining and wash buffers to block internalization. (Leave it out of live-cell functional assays.)
  • Keep cells cold from the first antibody step until fixation.
  • Fix promptly after the last wash.
  • If internalization or translocation is what you are measuring, control the timing deliberately: stain cold, warm for a defined time, then stop on ice and fix. Include a time-zero sample that was kept cold throughout.

Preparing cells

  • Cells and volume: at least 1 × 106 cells in 50 µl, which is 1–2 × 107 cells/ml. The exact concentration is not critical.
  • Tubes: 1.5 ml microcentrifuge tubes, siliconized (low-retention) if possible.
  • Buffer: PBS with 2% FBS (or 0.5–1% BSA), plus 0.1% sodium azide unless the cells must stay alive.
  • Single cells: pass the sample through a 70 µm strainer just before you bring it. If your cells clump, add EDTA or use Accumax before fixation. Aggregates spoil co-localization and morphology measurements.
  • Fixation: 2% paraformaldehyde, 20 min on ice, then wash and resuspend in buffer. If you permeabilize, fix first. This also meets the biosafety requirement for BSL-2 samples.
  • Viability: if you include dead cells, use a fixable viability dye.

Controls

  • Single-stain compensation controls for every fluorochrome. Use the same cells with the same fixation as your samples, not beads.
  • An unstained control.
  • Biological controls for the feature you measure, for example untreated vs. stimulated, or a known positive sample.

Biosafety

BSL-2 samples must be fixed before you bring them to the core: 2% paraformaldehyde, 20 min on ice, as in Preparing cells. The ImageStream is not in a biosafety cabinet. Samples that require containment above BSL-2 are not accepted.

Booking and the day of the run

  • First experiment: feasibility. Bring your single-stain controls and one or two samples. This tells us whether the staining and imaging will answer your question. Once feasibility is established, you can bring more samples in a single session.
  • Fill in the Data Acquisition Form (the button below, or at the top of this page) and paste its link into your iLab request. The form asks for your fluorochromes and the markers they detect, and for the image feature you want to measure (localization, co-localization, spot count, morphology, internalization…), which decides the magnification, event count and settings.
  • Bring samples on ice, protected from light, in clearly labelled tubes.
Required: the Data Acquisition Form Fill it in before your reservation. At the end, click Copy link and paste the link into your iLab request. We use it to set up the instrument for your experiment. https://www.cytochop.org/guidelines/imagestream-request.html
Open the Data Acquisition Form →

Reserve the ImageStream in iLab.

Your data and analysis

  • Acquisition software: INSPIRE®. It runs the instrument and collects the images. Its analysis tools are limited, so plan to analyse your data in IDEAS.
  • Files are large, often several GB per experiment. They include raw (.rif), compensated (.cif) and analysis (.daf) files. We share them through OneDrive, or we can copy them to your external hard drive if you bring one.
  • Analysis software: IDEAS® 6.2. It includes most analysis functions, including the analysis wizards. It does not include the machine-learning/AI modules.
    • Core workstations: IDEAS is installed on two core workstations, in ARC 1207i and MCRI 4410. They are free to use; reserve them in iLab.
    • Your own PC: IDEAS 6.2 is free; email flowcytometry@chop.edu for the download link. IDEAS runs on Windows only. It does not run on a Mac. Mac users can use the core workstations.